CHARACTERISTICS OF INSTALLATION
• optic unit;
• universal clamp;
• connecting cable;
• diaphragm;
• notebook;
• oscilloscope;
• kit of emission sources.
DESIGNED FOR
Reproducing, storing and transmitting of the unit of the average power and density power of the continuous and pulse-modu-lated optic emission, calibration and verification of integrated and autonomous pho-tometers for laser medical equipment and laser therapeutical, surgical and therapeu-tical equipment. Verification of measuring instruments can be carried out according to GOST R 8.706-2010.
SPHERES OF APPLICATION
Medicine: calibration and verification of integrated and autonomous photometers for laser medical equipment and laser therapeutical, surgical and therapeutical equipment.
| Range of measured values of average power, Wt | from 10-6 to 10 | |||||||
| Range of measured values of power density, Wt/m2 | from 10-1 to 106 | |||||||
| Range of wavelength of investigated emission, μm | from 0,45 to 1,1 | |||||||
| Calibration wavelength, fixed in ranges, μm | 0,6328; from 0,8 to 0,9 | |||||||
| Limits of permissible relative measurement error, %: | ||||||||
| • average power at calibration wavelength | ± 5,5 | |||||||
| • average power in working spectral range: | ||||||||
| – (from 0,6 to 1,0) μm | ± 6,5 | |||||||
| – (from 0,45 to 1,1) μm | ± 8,0 | |||||||
| • relative power levels | ± 4,0 | |||||||
| Rise time of transient response of measuring converter channel, ns,not more than | 15 | |||||||
| Limit of measuring converter linearity (on the load of 50 Om), V | 3 | |||||||
| Installation УПЛТ-М provides measurement of average power andpower density of the continuous and pulse-modulated emission: | ||||||||
| • in space angle, stere, up to | 2π | |||||||
| • at pulse repetition frequency, Hz, not less than | 50 | |||||||
| • at pulse duration, ns, not less than | 50 | |||||||
*Characteristics of installation УПЛТ -М can be specified by request of the customer.