The leak detector is a highly sensitive magnetic mass spectrometer configured to register a sample gas (helium) flow. The leakage is determined by the flow of helium entering the test volume during vacuum tests, or flowing out of the test volume at an overpressure in it.
DESIGNED FOR
Measuring helium flows during non-destructive testing of tightness.
Characteristics | МС-4 |
Range of indications at the leak detector input, Pa·m3/s | from 5•10-13 to 1·10-3 |
Measurement range at the leak detector input, Pa·m3/s | from 7•10-11 to 1·10-4 |
Range of indications when working with a probe method, Pa·m3/s | from 1•10-9 to 1 |
Limits of permissible relative measurement error , % from the measured value (where Qllm – lower limit of measurement, Qmfv – measured flow value, Pa·m3/s) | ±(0.50 + Qllm/Qmfv )·100 |
Overall dimensions, mm – primary pump in the leak detector housing – primary pump on a transport trolley |
605х400х401 545х645х1100 |
Weight, kg | 85 |
AC powered: – volume, V – frequancy, Hz |
220±22 50 |
Power consumption, V А | 800 |
Operating conditions: – ambient temperature, оС – relative humidity , % – atmospheric pressure of ambient air, kPa |
from +10 to +35 80 from 86 to 106.7 |
Average service life, years | 10 |