Secondary standard of unit of energy of pulse laser emission in рогtаblе design

DESIGNED FOR
Calibration and verification of measuring instruments of laser emission energy in their places of operation.

SPHERES OF APPLICATION
In medicine, in industry, construction, in scientific research, for energy measuring instruments, distribution of energy density, pulse duration and wavelength of laser emission in the wavelength range from 0.3 to 2.0 microns.

Nominal values of wavelength, µm 1.064; 0.532
Energy value range, J: at wavelength 1,064 µm 5∙10-3÷2∙10-1
Energy value range, J: at wavelength 0,532 µm 51∙0-3÷51∙0-2
Pulse duration, s (5-7)·10-9
Beam diameter at the standard outlet, mm 6–8
Time of measuring instrument verification at one wavelength, h 1
Comblned error of energy measurement, not more than 1.5∙10-2
Transfer error of unit of energy, not more than 1.0∙10-2
Comblned standard uncertainty of energy measurement,

not more than

3.0∙10-2