Milliteslameters ТП2-2У

Portable universal milliteslameter is designed for measuring magnetic induction of constant and variable (average rectified and peak value) magnetic field, as well as of single impulses of magnetic field in the range of 0.01 – 1999 mT.
It is convenient for studying magnetic systems of different purposes, in magnetic powder testing, for measurement of residual magnetization, for testing the magnetic therapeutic equipment of all kinds, and also for monitoring the magneticfields of various objects in a wide frequency range.

Sensitivity elements Hall`s sensors
Two measurement probes axial (type C and C1) and radial (type M and M1)
Measurable quantities • magnetic induction of constant magnetic field;
• magnetic induction of variable magnetic field with frequency from 20 Hz to 10 kHz (for modifications ТП2-2У, ТП2-2У-01, ТП2-2У-02);
• magnetic induction of impulse magnetic field with pulse duration from 100 µs to 5 s (for modifications ТП2-2У, ТП 2-2У-01, ТП2-2У-02).
Observation of the form of variable and pulsed magnetic field when connecting the
oscilloscope to the analog output is allowed.
Limits of measurement (Al) 20 mT, 200 mT, 2000 mT for modification ТП2-2У
Uncertainty ∆­₀, % when measuring constant magnetic field:
Ло = ± [ 2.0 + 0.1 (Al/Am – 1)]
when measuring variable magnetic field:
Ло = ± [ 2.5 + 0.2 (Al/Am – 1)]
when measuring impulse magnetic field:
Ло = ± [ 5.0 + 0.5 (Al/Am – 1)]

Milliteslameters ТП2-2У
Depending on the measured quantities and their measurement ranges, the milliteslameter
ТП2-2У produced in seven versions:

Version

Measured magnetic induction of magnetic field

Limits of measurement

constant

variable

impulse

ТП2-2У + + + 20, 200, 2000 mТ
ТП2-2У-01 + + + 2, 20, 200 mТ
ТП2-2У-02 + + + 200, 2000 mТ, 20 Т
ТП2-2У-03 + 20, 200, 2000 mТ
ТП2-2У-04 + 2, 20, 200 mТ
ТП2-2У-05 + 200, 2000 mТ, 20 Т
ТП2-2У-08 + 2000 mТ